The standard for emitting electromagnetic wave
As DIA:CONN P8 is manufactured to be used in the designated electromagnetic environment, it is used under the environment as below.
Test Name | Compliance Level | Electromagnetic Environment – Guidelines |
---|
RF emissions
CISPR 11
Group 1
As DIA:CONN P8 uses RF energy only for the inner function, the RF emission is very low.
There is little possibility of interfering with any adjacent electronic device.
RF emission
CIPR11
Class B
DIA:CONN P8 can be used in all living or other facilities directly conncted to a public power supply with low voltage,
which usually offers electricity to the home building.
High-frequency
wave emissions
IEC 61000-3-2
None
Voltage
Variation/flicker
emissions
IEC 61000-3-3
None
The standard for tolerance of electromagnetic wave
As DIA:CONN P8 is manufactured to be used in the electromagnetic environment, it should be used in the environment below.
Test Name | Test condition of IEC 60601 | Compliance Level | Electromagnetic Environment – Guidelines |
---|---|---|---|
Electrostatic discharge (ESD), IEC61000-4-2 | ± 8kV contact ± 2kV, ± 4kV, ± 8kV, ± 15kV stanby | ± 8kV contact ± 2kV, ± 4kV, ± 8kV, ± 15kV stanby | The floor is made of wood, concrete, or ceramic tile. If the floor is covered with synthetic material, the relative humidity should be more than 30%. |
Electronic high-speed transients/burst IEC 61000-4-4 | ± 1kV, ± 2kV @PRF 100kHz, meet BS /EP requirements | ± 1kV, ± 2kV @PRF 100kHz, meet BS /EP requirements | The quality of the main power should be commercial or in the hospital environment. |
Surge IEC 61000-4-5 | Common mode ± 0.5 kV, ± 1kV, ±2kV, Line-to-line (differential mode) ± 0.5 kV, ± 1kV | Common mode ± 0.5 kV, ±1kV, ±2kV, Line-to-line (differential mode) ± 0.5 kV, ± 1kV | The quality of the main power should be commercial or in the hospital environment. |
Voltage descent of power supply input line, Instant correction and voltage fluctuations IEC 61000-4-11 | 5s power interruption, meet BS requirements and meet EP requirements | <5% Uт 5 s power interruption, meet BS requirements and meet EP requirements | None |
Power frequency (50/60Hz) magnetic field IEC 61000-4-8 | 3A/m | 3A/m | The magnetic field of power frequency should be the representative level in commercial or hospital circumstances. |
Criteria for tolerance test of the outer connecting part of RF wireless telecommunication equipment
Test frequency MHz | Band aMHz | Service a) | Modulation | Level of tolerance test V/m |
---|
385
380~390
TETRA 400
Pulse modulation b)
18Hz
27
450
430~470
GMRS 460,
FRS 460
FM c)
±5kHz deviation
1 kHz sine
28
710
704~787
LTE band 13,17
Pulse modulation b)
217Hz
9
745
780
810
800~960
GMS 800/900,
TETRA 800,
iDEN 830,
CDMA 850,
LTE band 5
Pulse modulation b)
18Hz
28
870
930
1720
1700~1990
GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE band
1, 3, 4, 25; UMTS
Pulse modulation b)
217Hz
28
1845
1970
2450
2400~2570
Bluetooth, WLAN,
802.11 b/g/n, RFID 2450,
LTE band 7
Pulse modulation b)
217Hz
28
5240
5100~5800
WLAN 802.11 a/n
Pulse modulation b)
217Hz
9
5500
5785
To reach the level of tolerance test, if needed, the distance between the transmission antenna and the ME device or ME system may be decreased to 1m. The test distance of 1m can be accepted by IEC 61000-4-3. |
a) Some services include only uplink frequency. b) Carrier wave should be modulated using square wave signal with 50% duty cycle. c) As an alternative for FM modulation, the carrier wave, can modulate the pulse using a 50% duty square wave signal at 18Hz. This does not mean the real modulation but is the worst case. |